Updates :
WELCOME TO IACTA ECHO LIBRARY           
      




Mitral Valve Area by Pressure Half-Time (PHT)


Calculator

PHT
msec

Mitral Valve Area (MVA)
cm²
Aortic Valve Gradient

How to get an MVA by PHT.

Step 1: Obtain a PWD or a CWD of the mitral valve. The flow profile of the mitral valve will have a Vmax, typically the maximum velocity of the E wave. The time from the Vmax to the velocity equal to Vmax divided by 1.4 is the pressure half time. (Vmax / 1.4 is equivalent to the half pressure). 220 divided by the PHT is the mitral valve area.